07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 9-9
9. Operating Instructions – Change Program
9.7 Standard Mode (Standard Fill Mode)
The following sections define the therapy settings for Standard Fill Mode. For Low
Fill Mode settings, see 9.8, Low Fill Mode, on page 9-34.
9.7.1 CCPD/IPD Therapy Settings
Table 9-3 shows the settings programmed for CCPD/IPD therapy in the Standard
Fill Mode.
Table 9-3. CCPD/IPD Settings – Standard Fill Mode
Setting Description
Total Volume
Total Volume of solution used for the therapy. Includes the
total Fill Volume for all cycles and the Last Fill Volume.
Minimum setting is 200 mL
Maximum setting is 80000 mL
Default setting is 200 mL
Setting increments:
– 50 in the 200 to 2000 mL range
– 100 in the 2000 to 5000 mL range
– 500 in the 5000 to 80000 mL range
Therapy Time
Total time for the nighttime portion of therapy. This time is
fixed and begins with Initial Drain.
Minimum setting is 10 minutes
Maximum setting is 48 hours
Default setting is 10 minutes
Setting increment is 10 minutes