07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 11-11
11. Operating Instructions – Prepare for Therapy
Follow the steps below to load the disposable set.
WARNING
Before loading the disposable set, inspect the cassette and
tubing for damage. Using damaged sets can result in
contamination of the fluid or fluid pathways.
Contamination of any portion of the fluid or fluid path can
result in peritonitis.
Inspect the flexible surfaces of the cassette for obvious
signs of damage, including cuts, tears, or punctures.
Ensure the tip protectors on the ends of the tubing are
on and unbroken.
If damage is found, obtain a new disposable set and repeat
inspection procedure.
Tubing indentations can be present on disposable sets due
to the supple nature of the tubing. Slight tubing
indentations are cosmetic in nature and should have no
impact on the functionality of the product.
Steps to load the disposable set
1. Prepare the disposable set.
Open the packaging and remove the
disposable set.
Close all clamps.
– 3-prong set has 5 clamps
– 4-prong set has 6 clamps