18. Troubleshooting
18-50 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
Steps to bypass
Caution: Negative UF alarm Display screen
CAUTION: NEGATIVE UF appears on the display screen.
1. Press
STOP to silence the alarm.
The CAUTION: NEGATIVE UF alarm alternates with
the current DRAIN phase.
alternates with
h NOTE: The default value for the Negative UF alarm is 50% of your
prescribed Fill Volume.
h NOTE: The Negative UF alarm limit can be programmed at 20% to 60% in
the Low Fill Mode.
2. Press
V.
h NOTE: The amount that your Drain Volume is short of your prescribed Fill
Volume is subtracted from the previous Drain Total UF (see below) to
obtain your current UF that resulted in the Negative UF alarm.
3. Press
V.
The Initial Drain Volume (I-DRAIN VOL) from the
current therapy is shown.
4. Press
V.
The TOTAL UF from the end of the previous Drain is
shown.
5. Press
V.
h NOTE: Unless you have absorbed a lot of fluid, resume the Drain or select
MANUAL DRAIN. Using Manual Drain repeatedly does not generate
an audible alarm.
CAUTION: NEGATIVE UF
CAUTION: NEGATIVE UF
DRAIN 2 OF 5
DRAIN VOLUME: ML
I-DRAIN VOL: ML
TOTAL UF: ML
BYPASS