18. Troubleshooting
18-42 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
Steps to bypass a Drain phase Display screen
DRAIN phase number appears on the display screen.
1. Press
STOP.
2. Press V.
The DRAIN VOLUME is shown.
3. Press
V.
The Initial Drain Volume (I-DRAIN VOL) is shown.
4. Press
V.
The TOTAL UF at the end of your previous Drain is
shown.
5. Press
V.
6. Press
ENTER.
The next FILL will begin. It will be a full Fill if no
alarm occurred.
– OR –
If the volume drained is less than your Minimum
Drain Volume, a DRAIN NOT FINISHED alarm
appears if you try to bypass. See 18.4.4, Bypass
Drain Not Finished Alarm, on page 18-43.
– OR –
A CAUTION: NEGATIVE UF alarm may appear. See
18.3.11, Caution: Negative UF, on page 18-22 and
18.4.6, Bypass Caution: Negative UF Alarm, on
page 18-49.