07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 18-19
18. Troubleshooting
18.3.8 Slow Flow Rate
SLOW FLOW
Display Message:
Cause: Flow rate is very slow. A slow flow rate can reduce the Dwell
Time and decrease the amount of effective dialysis time.
There can be a partial kink on the specified line.
This is an Auto Restart Alarm.
To Correct: 1. Check the line shown in the display for:
Kinks
Closed clamps
Fibrin blockage
Disconnected solution bags
Empty solution bags
2. Correct the problem you found.
You do not need to press any buttons.
– OR –
If the alarm becomes a Manual Restart Alarm:
1. Press
STOP to mute the alarm.
2. Check the line shown in the display for:
Kinks
Closed clamps
Fibrin blockage
Disconnected solution bags
Empty solution bags
3. Correct the problem you found.
4. Press
GO to return to the therapy.
SLOW FLOW DRAIN
SLOW FLOW HEATER
SLOW FLOW PATIENT
SLOW FLOW SUPPLY