
07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 18-53
18. Troubleshooting
18.5 Manual Drain Procedure
WARNING
Do not STOP or BYPASS a Manual Drain during Fill. An
increased intraperitoneal volume (IIPV) situation can
result.
IIPV can result in a feeling of abdominal discomfort,
serious injury, or death.
If any patient, or patient caregiver, suspects the patient has
IIPV during a treatment, follow the Manual Drain
procedure below. See 18.8, Increased Intraperitoneal
Volume (IIPV), on page 18-58 for more information.
Additional care should be taken to monitor for IIPV
symptoms for those patients not able to communicate
essential information to their caregiver during treatment.
Steps to perform a Manual Drain Display screen
The current FILL phase appears on the display screen.
1. Press
STOP.
2. Press V.
3. Press
V.
4. Press
V.
5. Press
V.
6. Press
V.
Comentários a estes Manuais