07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 13-9
13. Operating Instructions – End Therapy
Steps to shut down the cycler
h NOTE: Close all clamps and disconnect yourself by following the steps in
13.2, Disconnect Yourself, on page 13-6 before continuing.
1. Press
GO.
The display alternates between CLOSE ALL
CLAMPS and DISCONNECT YOURSELF.
The occluder (behind door) retracts for 2 minutes
(30 seconds for Low Fill Mode) to allow removal of
the cassette.
2. Lift the latch up to unlock and open door.
3. If the occluder closes before you open the door,
press
STOP and then GO.
The occluder retracts again for a short time.
4. Remove and discard the disposable set and solution
bags.
Follow your local guidelines for disposal of dialysis
waste materials.