07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 12-15
12. Operating Instructions – Perform Therapy
12.6.1 Day Dwell Options
Table 12-6 lists the options you can select during the DAY DWELL phase in
Hi-Dose Therapy.
1. Press
V to view each option.
2. Press
ENTER to select an option.
3. Press
STOP to return to the previous menu.
Table 12-6. Options Available During Day Dwell
Option Description
Day Dwell Time
The amount of time elapsed during the daytime Dwell.
Continue
Therapy
Connect
Yourself
Press
GO twice to drain the Day Fill and perform the next
cycle.
Initial Drain
Volume
Volume from Initial Drain.
Total
Ultrafiltration
The total therapy UF, updated after each Drain is
completed. Press
ENTER to review cycle-by-cycle
information. Press
STOP to return to the previous menu.
DAY DWELL X OF X
DAY DWELL TIME: 0:31
PRESS GO TO CONTINUE
CONNECT YOURSELF
I-DRAIN VOL: 65ML
TOTAL UF: 252ML