Red Cycler 5C Instruções de Operação Página 302

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 301
07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 18-47
18. Troubleshooting
5. Press V.
Bypassing confirms that you are empty and that
you want the next Fill delivery to equal your
prescribed Fill Volume.
You will achieve negative UF if you bypass the Low
Drain Alarm and your Drain Volume is less than
your Fill Volume. Your UF decreases by the amount
your Drain is short of your prescribed Fill Volume.
6. Press
GO to return to Drain.
This is the recommended option.
– OR –
Press
ENTER to bypass.
The next FILL begins. Your full Fill Volume is
delivered.
If you pressed
ENTER and the increase in negative
UF exceeds your Negative UF Limit, a CAUTION:
NEGATIVE UF alarm occurs. See 18.4.6, Bypass
Caution: Negative UF Alarm, on page 18-49.
a. Press
STOP to silence the alarm.
alternates with
b. Press GO to resume DRAIN.
h NOTE: Change your position after you resume Drain. Occasionally, the
location of the catheter tip can be in a less-than-optimal position.
Changing your position may resolve the low Drain flow issue.
Steps to bypass
Low Drain Volume alarm (Continued) Display screen
BYPASS
DRAIN 2 OF 5
FILL 3 OF 5
CAUTION: NEGATIVE UF
CAUTION: NEGATIVE UF
DRAIN 2 OF 5
DRAIN 2 OF 5
Vista de página 301
1 2 ... 297 298 299 300 301 302 303 304 305 306 307 ... 376 377

Comentários a estes Manuais

Sem comentários