07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 9-57
9. Operating Instructions – Change Program
WARNING
Your Day Fill Volume in milliliters (mL) should normally not exceed
the values shown in Table 19-7 on page 19-15. Contact your dialysis
center to confirm your Day Fill Volume if it exceeds this volume.
Exceeding this volume can result in an increased intraperitoneal
volume (IIPV) situation.
IIPV could result in a feeling of abdominal discomfort, serious injury,
or death. See page 18-58 for IIPV symptoms or if IIPV is suspected.
Additional care should be taken to monitor for IIPV symptoms for
those patients not able to communicate essential information to their
caregiver during treatment.
Day Fill Volume
Volume of solution for each daytime exchange. This
parameter appears only when the Number of Day Fills has
been set to greater than zero (0).
Minimum setting is 60 mL
Maximum setting is 1000 mL
Default setting is 250 mL
Setting increments:
1 in the 60 to 100 mL range
10 in the 100 to 500 mL range
50 in the 500 to 1000 mL range
Night Therapy
Time
Total time for the nighttime portion of therapy. This time
begins as soon as you start the last Drain of the daytime
exchange.
Minimum setting is 10 minutes
Maximum setting is 48 hours
Default setting is 10 minutes
Setting increment is 10 minutes
Table 9-17. Hi-Dose Tidal Settings – Low Fill Mode (Continued)
Setting Description
DAY FILL VOL: ML
NITE THER TIME:HH:MM