07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 18-51
18. Troubleshooting
6. Press ENTER to bypass.
alternates with
h NOTE: Change your position to assist draining. Fluid may have pocketed
near your catheter.
7. Press
GO to resume DRAIN.
– OR –
a. Press
V until MANUAL DRAIN appears.
b. Press
ENTER.
– OR –
a. Press
ENTER to bypass.
The next FILL begins.
h NOTE: By selecting BYPASS, you indicate that you
are empty. The system considers your
volume zero (0) and delivers your entire
prescribed Fill Volume.
h NOTE: Bypassing a CAUTION: NEGATIVE UF alarm will temporarily increase
your Minimum Drain Volume percentage to 100% of your prescribed
Fill Volume. If your Drain Volume percentage is greater than 100%, it
stays the same. Additional bypassing is allowed only if this Minimum
Drain Volume has been met. Your Minimum Drain Volume
percentage will remain at 100% until your UF falls below the
Negative UF alarm limit (default value is 50%).
Steps to bypass
Caution: Negative UF alarm (Continued) Display screen