18. Troubleshooting
18-10 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
18.3.3 Check Therapy Setting Value
Check value
Display Message:
Cause: The value of the programmed therapy setting is not valid. If
you are using a HomeChoice/HomeChoice PRO APD System
for the first time, this alarm occurs to remind you to program
your therapy.
This is a Manual Restart Alarm.
To Correct: 1. Press
STOP to mute the alarm.
The display screen automatically returns to the
incorrect setting and flashes the value you need to
change.
2. Change the value or other values.
3. Press
STOP to exit Change Program.
CHECK THERAPY TIME
CHECK TIDAL VOL PCT
CHECK TOTAL UF
CHECK TOTAL VOLUME