13. Operating Instructions – End Therapy
13-8 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
13.3 Shut Down
WARNING
Discard the disposable set and all solution bags at the end
of therapy. Possible contamination of the fluid or fluid
pathways could result if disposables are reused.
Contamination of any portion of the fluid or fluid path may
result in peritonitis.
WARNING
Do not open the door until you have CLOSED ALL CLAMPS.
This prevents the flow of fluid from one bag to another,
and/or to the patient. Uncontrolled gravity flow of fluid can
result in an increased intraperitoneal volume (IIPV)
situation.
IIPV can result in a feeling of abdominal discomfort,
serious injury, or death.
If any patient, or patient caregiver, suspects the patient has
IIPV during a treatment, press
STOP immediately, then
press
V and initiate a Manual Drain. The Manual Drain
procedure is located in 18.5, Manual Drain Procedure, on
page 18-53. See 18.8, Increased Intraperitoneal Volume
(IIPV), on page 18-58 if IIPV is suspected.
Additional care should be taken to monitor for IIPV
symptoms those patients not able to communicate
essential information to their caregiver during treatment.