11. Operating Instructions – Prepare for Therapy
11-8 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
Initial Drain
Volume
Shows the volume from the Initial Drain of your last
treatment.
Last Manual
Drain
Only appears if you drained some fluid using the Manual
Drain option after a Last Fill. Shows the amount of solution
drained.
Last
Ultrafiltration
Amount of ultrafiltrate removed during the last treatment.
Press
ENTER to review cycle-by-cycle information. Press
STOP to return to the previous menu.
If this value is lower than usual, or negative, temporarily
increase your I-Drain Alarm setting when Verify I-Drain is
displayed prior to the start of the Initial Drain. This ensures
a complete Initial Drain. See 10.2.7, I-Drain Alarm, on
page 10-9.
Average Dwell
Time
Average actual Dwell Time per cycle for the last treatment.
Press
ENTER to review cycle-by-cycle information. Press
STOP to return to the previous menu.
Alarm Log
Press
ENTER to review the last 20 alarms. Press STOP to
return to the previous menu.
LOG IS EMPTY appears if no alarms occurred.
Table 11-1. Options Available at Startup (Continued)
Option Description