07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 18-35
18. Troubleshooting
18.3.18 Temp Stabilizing
TEMP STABILIZING
Message Display:
Cause: The temperature of the fluid in the heater bag measured by
the cycler is above 40°C.
Fluid delivery is not allowed at this time.
The room temperature combined with the heat from the
system (even with the bag heater off) is raising the solution
bag temperature too high to deliver safe fluids to you.
To Correct: 1. Press the power switch OFF.
2. Allow the system to cool for 10 to 20 minutes.
Bags containing less fluid cool in less time.
3. Press the power switch ON.
POWER RESTORED appears on the display screen.
4. Press
STOP.
5. Press
GO.
Therapy automatically resumes if the fluid temperature
falls below 40°C.
If the TEMP STABILIZING message continues to appear:
1. Press the power switch OFF.
2. Wait another 20 minutes.
3. Repeat Steps 3 through 5 above.