07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 19-5
19. Technical Data
Table 19-1. Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The HomeChoice/HomeChoice PRO APD System is intended for use in the electromagnetic environment specified below.
The user of the system should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment – Guidance
RF emissions CISPR 11 Group 1 The HomeChoice/HomeChoice PRO APD System uses RF energy
only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11 Class B The HomeChoice/HomeChoice PRO APD System is suitable for use
in all establishments, including domestic establishments and those
directly connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Complies