Red Cycler 5C Instruções de Operação Página 287

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 286
18. Troubleshooting
18-32 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
SYSTEM ERROR 2240 or 2267
Message Display:
Cause: The system detected air in the disposable tubing set and
solution bag, and diverted flow to drain.
SYSTEM ERROR 2240 and SYSTEM ERROR 2267 can be
caused by one or more of the following:
Leaks
Disconnected disposable tubing
Loose connections
Fluid level not at or near the patient connector after
completing the prime cycle (incomplete prime)
Unclamped, unused supply lines (if the tip protector has
been disrupted)
Use of “Dummy Tummy” during training
To Correct: 1. Press
STOP to mute the alarm.
2. Write down the System Error number and treatment
phase (Fill, Drain, or Dwell) that appears on the display
screen.
3. Turn the power switch OFF and ON to end therapy.
SYSTEM ERROR 2367 appears.
4. Press the power switch OFF and ON again.
PRESS GO TO START appears.
5. Close all clamps.
6. Discard the disposable set and solution bags.
7. Notify your dialysis nurse.
h NOTE: If you need help to correct the alarm, contact Baxter Technical
Assistance at the number listed in 2.2, Numbers to Call for Assistance,
on page 2-1.
SYSTEM ERROR 2240
SYSTEM ERROR 2267
Vista de página 286
1 2 ... 282 283 284 285 286 287 288 289 290 291 292 ... 376 377

Comentários a estes Manuais

Sem comentários