18. Troubleshooting
18-28 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
Table 18-1. Possible Causes for Reload the Set Alarms
Reload Set Number Possible Cause
143, 163, 165–169 The occluder is unable to pinch off flow due to:
Cold solution
Cassette tubing overlapping or touching where it
exits the door
h NOTE: Placing the heater bag and cassette on the
heater pan prior to the start of therapy can
often prevent this alarm.
134–137, 156, 157 Clamped or kinked line
WARNING
To prevent delivery of non-sterile air to your peritoneal cavity, load a
new set and bags if RELOAD SET: 201 occurs and fluid flows from the
patient line. Non-sterile air in the peritoneal cavity can cause
peritonitis. Air in the peritoneal cavity can cause abdominal and/or
shoulder pain.
201
Pneumatic leak between cassette and membrane
gasket due to incorrect loading
Debris on outside of cassette
Hole in cassette sheeting
h NOTE: Load a new set and bags if RELOAD SET: 201
occurs and fluid flows from the patient line.
h NOTE: Always check cassette for debris, slits, tears,
or punctures before use.
200, 202, 203
Pneumatic leak between cassette and membrane
gasket due to incorrect loading
Debris on outside of cassette
Hole in cassette sheeting
h NOTE: Always check cassette for debris, slits, tears,
or punctures before use.
All others Incorrect loading of set or issue with cassette