07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 9-31
9. Operating Instructions – Change Program
h NOTES: (Continued)
Seventy percent (70%) of your normal Night UF is a
good starting point for determining your optimum
Total UF. For help in converting 70% of your
expected total therapy UF into a value that can be
programmed as your Total UF for your Tidal
therapy, see 19.17, Determining Tidal Total UF and
Last Manual Drain UF Target Volume Settings, on
page 19-18.
If you use a solution for your Tidal therapy that is
different from the solution used in your previous
therapy, you may need to adjust your Total UF based
on the concentration of the new solution. Contact
your dialysis center for recommendations regarding
setting your Total UF in this situation.
Last Fill
Volume
Last Fill Volume delivered at the end of therapy and left in the
peritoneal cavity during the day. Also called “Wet Day.”
Minimum setting is 0 or 100 mL
Maximum setting is 3000 mL
Default setting is 0 mL
Setting increments:
10 in the 100 to 500 mL range
50 in the 500 to 1000 mL range
100 in the 1000 to 3000 mL range
Table 9-9. Hi-Dose Tidal Settings – Standard Fill Mode (Continued)
Setting Description
LAST FILL VOL: ML