18. Troubleshooting
18-46 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
2. Press V.
The DRAIN VOLUME is shown.
3. Press
V.
The Initial Drain Volume (I-DRAIN VOL) from the
current therapy is shown.
4. Press
V.
The TOTAL UF from the end of the previous Drain
is shown.
WARNING
Bypassing a LOW DRAIN VOLUME alarm when the Total UF at the end
of the last cycle is lower than normal for this point in the therapy can
result in an increased intraperitoneal volume (IIPV) situation. A
negative UF value can result in a greater potential for IIPV. Press
GO
to resume Drain.
IIPV can result in a feeling of abdominal discomfort, serious injury, or
death.
If any patient, or patient caregiver, suspects the patient has IIPV
during a treatment, press
STOP immediately, then press V and
initiate a Manual Drain. The Manual Drain procedure is located in
18.5, Manual Drain Procedure, on page 18-53. See 18.8, Increased
Intraperitoneal Volume (IIPV), on page 18-58 if IIPV is suspected.
Additional care should be taken to monitor for IIPV symptoms for
those patients not able to communicate essential information to their
caregiver during treatment.
Steps to bypass
Low Drain Volume alarm (Continued) Display screen