Red Cycler 5C Instruções de Operação Página 267

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 266
18. Troubleshooting
18-12 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
18.3.5 Load a New Set
To Correct: 1. Press STOP to mute the alarm.
2. Check with your dialysis center to learn when it is safe
to bypass. See 18.4, Bypass Procedures, on page 18-37.
3. If you are sure that you want to bypass the alarm or
phase, press
V until the display screen shows BYPASS.
4. Press
ENTER to select BYPASS.
You have bypassed the alarm or phase and moved to the
next phase in your therapy.
LOAD A NEW SET
Display Message:
Cause: The disposable set has failed testing during setup.
This is a Manual Restart Alarm.
To Correct: 1. Press
STOP to mute the alarm.
2. Remove and discard the disposable set, and solution
bags if necessary.
3. Get a new disposable set, and new solution bags if
needed.
4. Load the cassette.
5. Press
GO and follow the setup instructions on the
display screen.
NOT FINISHED (Continued)
LOAD A NEW SET
LOAD NEW SET & BAGS
Vista de página 266
1 2 ... 262 263 264 265 266 267 268 269 270 271 272 ... 376 377

Comentários a estes Manuais

Sem comentários