3. Warnings and Cautions
3-14 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
3.3 Cautions
The cycler you are using may be the property of Baxter Healthcare
Corporation. Improper care or use may result in additional expense.
Wipe up any spills right away. This reduces the chance of moisture
entering the system and causing a malfunction. This also reduces the
chance of bacteria contamination or an unsanitary condition.
To prevent the cycler from falling, place it on a sturdy, stable nightstand
or table large enough to hold the cycler and the solution bags. Falling can
damage the cycler or cause personal injury.
If an extension cord is used, make sure the ampere rating of the
HomeChoice/HomeChoice PRO APD System does not exceed the extension
cord ampere rating.
– Use only heavy-duty extension cords rated at 1200 watts (10 amp for
110V systems).
– Use only extension cords less than 12 feet (3.5 meters) in length.
– The extension cord must have the third grounding wire that mates
with the grounding plug on the HomeChoice/HomeChoice PRO APD
System power cord.
Failure to follow this advice can result in excessive heating or fire.
Do not attempt to service the HomeChoice/HomeChoice PRO APD System
yourself. Doing so can result in fire, burns, electrocution, or other
personal injury. For servicing, contact Technical Assistance at the number
listed in 2.2, Numbers to Call for Assistance, on page 2-1. Baxter Technical
Assistance is available 24 hours a day for HomeChoice/HomeChoice PRO
APD System users. A service manual for this product is not available.