07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 18-5
18. Troubleshooting
18.3 Alarms
18.3.1 Check Lines
WARNING
Do not replace empty solution bags or reconnect
disconnected solution bags during your therapy.
Possible contamination of the fluid or fluid pathways can
result if disposables are reused. Contamination of any
portion of the fluid or fluid path can result in peritonitis.
If a bag becomes disconnected during your therapy, follow
the End Therapy Early procedure. Discard the disposable
set and all solution bags at the end of therapy. (See
18.6, End Therapy Early Procedure, on page 18-55.)
CHECK LINE
Display Message:
Cause 1: This is an Auto Restart Alarm.
The line referenced on the display screen is blocked due to:
– Kinks
– Closed clamps
– Fibrin blockage
– Disconnected solution bags
– Empty solution bags