9. Operating Instructions – Change Program
9-46 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
Table 9-15 shows the settings programmed for Tidal therapy in the Low Fill Mode.
WARNING
Patients whose Fill volumes are less than 1000 mL may
normally drain slowly. These patients typically weigh less
than 44 lbs (20 kg). Use of the Low Fill Mode minimizes the
incidence of LOW DRAIN VOLUME and CAUTION: NEGATIVE
UF alarms. These alarms can not be bypassed to prevent an
increased intraperitoneal volume (IIPV) situation in Low
Fill Mode. The Negative UF Limit should not be raised
above 50% and the Minimum Drain Volume % should not
be lowered below 85%, (the default values). It is required
that the Low Recirculation Volume Set be used with Low
Fill Mode procedures.
IIPV could result in a feeling of abdominal discomfort,
serious injury, or death. See page 18-58 for IIPV symptoms
or if IIPV is suspected.
Additional care should be taken to monitor for IIPV
symptoms for those patients not able to communicate
essential information to their caregiver during treatment.
h
NOTE:
The Tidal Volume (Fill Volume x Tidal Volume %) must be at least
95 mL or a CHECK TIDAL VOL PCT alarm will occur. See 18.3.3, Check
Therapy Setting Value, on page 18-10.
h
NOTE:
Prescription settings can not be adjusted during a Tidal therapy.