18. Troubleshooting
18-26 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
18.3.13 Reload the Set
To Correct
Cause 2:
If you suspect this is the cause:
1. Press
STOP.
2. Press V until you see MANUAL DRAIN.
3. Press
ENTER to initiate a Manual Drain. Then open the
clamp on the patient line.
Do not open the clamp until after you press
ENTER.
4. When the Manual Drain ends, STOPPED: FILL appears.
5. Press
GO to resume Fill.
WARNING
Do not open the door until you have CLOSED ALL CLAMPS.
This prevents the flow of fluid from one bag to another,
and/or to the patient. Uncontrolled gravity flow of fluid can
result in an increased intraperitoneal volume (IIPV)
situation.
IIPV can result in a feeling of abdominal discomfort,
serious injury, or death.
If any patient, or patient caregiver, suspects the patient has
IIPV during a treatment, press
STOP immediately, then
press
V and initiate a Manual Drain. The Manual Drain
procedure is located in 18.5, Manual Drain Procedure, on
page 18-53. See 18.8, Increased Intraperitoneal Volume
(IIPV), on page 18-58 if IIPV is suspected.
Additional care should be taken to monitor for IIPV
symptoms for those patients not able to communicate
essential information to their caregiver during treatment.
CHECK YOUR POSITION (Continued)