10. Operating Instructions – Make Adjustments
10-14 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
10.2.9 Last Manual Drain
Baxter recommends YES be selected for the LAST MANUAL DRAIN option. A good
starting point for setting your Target UF is 70% of your expected Total UF. If the UF
is less than the Target UF at the end of the last Drain of the therapy, the system
stops and a LOW UF alarm appears on the display screen.
Occasionally, the location of the catheter tip can be in a less-than-optimal position.
This can lead to an incomplete Drain of the dialysis solution when lying down.
With the Last Manual Drain option, you may want to change position before the
Last Fill is performed by the system.
Follow the steps below to set Last Manual Drain.
h
NOTE:
If a LAST MANUAL DRAIN is set to YES, a UF Target and a UF Alarm
must be set. See 10.2.10, UF Target and Alarm, on page 10-15.
Steps to set Last Manual Drain Display screen
1. Press
ENTER to access the MAKE ADJUSTMENTS
menu.
2. Press
V until LAST MANUAL DRAIN appears.
3. Press
ENTER.
The N (NO) or Y (YES) blinks.
(N or Y blinks)
4. Press U or V to change the setting.
5. Press
ENTER to save the new setting.
The Y or N stops blinking.
(Blinking stops)
6. If LAST MANUAL DRAIN is set to YES, press V.
UF TARGET appears.
– OR –
7. If LAST MANUAL DRAIN is set to NO, press
STOP
to exit Make Adjustments.
MAKE ADJUSTMENTS
LAST MANUAL DRAIN: N
LAST MANUAL DRAIN: N
LAST MANUAL DRAIN: Y
LAST MANUAL DRAIN: Y
UF TARGET: 0ML
PRESS GO TO START