18. Troubleshooting
18-24 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
18.3.12 Check Your Position
– OR –
a. Press
V until MANUAL DRAIN appears.
b. Press
ENTER.
– OR –
a. Press
ENTER to bypass the alarm.
CAUTION: NEGATIVE UF appears briefly, then
the next FILL begins.
h NOTE: By selecting BYPASS, you indicate that you are empty. The system
considers your volume zero (0) and delivers your entire prescribed
Fill Volume.
CHECK YOUR POSITION
Message Display:
Cause 1: Your position is more than one foot (12 inches) above the
cycler.
This is an Auto Restart Alarm.
To Correct
Cause 1:
1. Check the position of the system.
2. If the system is too low, raise it by at least six (6) inches.
You do not need to press any buttons.
– OR –
Steps to check Total UF (Continued) Display screen