18. Troubleshooting
18-38 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
3. Press V.
4. Select one of the following options:
a. Press
GO to return to INITIAL DRAIN if you do
not want to bypass.
This is the recommended option.
h NOTE: Unless you have absorbed a lot of fluid, resume the Drain or select
MANUAL DRAIN. Using Manual Drain repeatedly does not generate
an audible alarm.
h NOTE: The Manual Drain can facilitate obtaining a complete Initial Drain. If a
Manual Drain ends due to Slow/No Flow, the system will revert to
STOPPED: DRAIN without alarming.
– OR –
a. Press
V until MANUAL DRAIN appears.
b. Press
ENTER.
– OR –
a. Press
ENTER.
Initial Drain is bypassed and the first FILL
begins.
– OR –
An alarm sounds if less than your Initial Drain
Alarm limit has been drained.
See 18.4.4, Bypass Drain Not Finished Alarm, on
page 18-43.
alternates with
Steps to bypass Initial Drain (Continued) Display screen