07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 19-7
19. Technical Data
Table 19-3. Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The HomeChoice/HomeChoice PRO APD System is intended for use in the electromagnetic environment specified below.
The user of the system should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level Electromagnetic Environment – Guidance
Portable and mobile RF communications equipment should be used no closer to
any part of the
HomeChoice/HomeChoice PRO APD System, including cables,
than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance:
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3Vrms 150 kHz
to 80 MHz
3V/m 80 MHz
to 2.5 GHz
3V
3V/m
d = [3.5 / 3]
d = [3.5 / 3] 80 MHz to 800 MHz
d = [7 / 3] 800MHz to 2.5GHz
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer, and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
a
, should be less than the compliance level
in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the
following symbol:
NOTES:
At 80 MHz and 800 MHz, the higher frequency applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast, and TV broadcast can not be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the system is used exceeds the applicable RF compliance level above, the system should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the system.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.