9. Operating Instructions – Change Program
9-62 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
9.8.8 Hi-Dose Tidal Calculated Settings
The HomeChoice/HomeChoice PRO APD System calculates the number of night
cycles and the Dwell Time. For a Tidal therapy, the system also calculates Tidal
Volume and ultrafiltration (UF) per cycle. The calculated values are shown on the
display screen when you press
STOP after you have completed reviewing your
therapy settings.
Table 9-18. Definitions for Hi-Dose Tidal Calculated Settings
Setting Description
Cycles Total number of cycles at night, not including the Last Fill.
This parameter is calculated by the system.
Dwell Time Calculated amount of time the dialysis solution remains in
the peritoneal cavity during each cycle. The system may
automatically adjust the Dwell Time based on your actual
flow rates during Fill and Drain, if your clinician chooses this
option.
Tidal Volume Actual Tidal Volume calculated based on the Tidal Volume %
programmed and the Night Fill Volume.
UF Per Cycle Estimated UF Per Cycle based on Total UF programmed and
number of cycles calculated.