07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 19-3
19. Technical Data
19.5 Battery Backup
When power failure occurs during therapy, the HomeChoice/HomeChoice PRO APD
System stops the therapy and the display screen turns off. However, the display
powers up each time a key is pushed during the first 30 minutes of a power failure.
If power is restored within two hours, the therapy can be restarted from where it
left off. If the power is not restored within two hours, the therapy is not allowed to
continue.
19.6 Electromagnetic Compatibility
The HomeChoice/HomeChoice PRO APD System, as with all medical electrical
equipment, needs special precautions regarding EMC (electromagnetic
compatibility), and the following information must be followed when installing
and putting the system into service.
Because the intensity of electromagnetic energy is greatest near the source of a
transmitting antenna, portable and mobile RF communications equipment can
affect medical electrical equipment.
The HomeChoice/HomeChoice PRO APD System has been designed to withstand
the effects of EMI (electromagnetic interference) and meets the most current EMC
standards that apply to the cycler. However, extremely high levels of
electromagnetic energy (above the levels of IEC 60601-1-2) may still produce
interference.
To reduce the risk of EMI, follow these recommendations:
Do not turn on or use hand-held personal communications devices, such
as mobile two-way radios or cellular phones, near the cycler. If these
devices need to be used, follow the recommended separation distance as
shown in the following tables.
ONLY DEVICES MARKED WITH THIS SYMBOL ARE
COMPLIANT TO THE IEC 60601-1-2 STANDARD
C
US
R