07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 8-11
8. Operating Instructions – PRO Card and Modem
8.3 Display Messages
The PRO Card port has an indicator light located on the control panel directly
above the port. See Figure 8-1 on page 8-2. This light can be green or yellow.
Green means the card is fully inserted and working.
Yellow means there is a problem with the PRO Card or card reader inside
the cycler.
If there is a problem with the PRO Card or the card reader, the following messages
can appear.
8.3.1 No PRO Card
Display Message:
Indicator Light: Flashing Yellow
Cause: The PRO Card is not inserted when you turn on the device, or
at the end of your therapy.
To Correct: 1. Turn the power switch OFF.
2. Insert the PRO Card and turn the power switch ON.
The indicator light turns Green if the PRO Card is
inserted correctly.
– OR –
1. If you can not correct the problem, press
GO to
continue.
No therapy information is recorded to the PRO Card.
You must report this to your dialysis center.