07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 9-59
9. Operating Instructions – Change Program
WARNING
A Total UF volume set too low can result in a gradual buildup of UF
volume during the therapy. This can result in an increased
intraperitoneal volume (IIPV) situation.
IIPV could result in a feeling of abdominal discomfort, serious injury,
or death. See page 18-58 for IIPV symptoms or if IIPV is suspected.
Additional care should be taken to monitor for IIPV symptoms for
those patients not able to communicate essential information to their
caregiver during treatment.
Night UF
Total ultrafiltration (UF) expected for the therapy. The
system calculates the UF Per Cycle. The UF Per Cycle plus the
Tidal Volume is the amount of solution drained during each
Tidal Drain.
Minimum setting is 0 mL
Maximum setting is 10000 mL
Default setting is 0 mL
Setting increments:
– 10 in the 0 to 1000 mL range
– 100 in the 1000 to 10000 mL range
h NOTES:
When the Therapy type is changed from CCPD to
Tidal, the Night UF reverts to the default setting of
zero (0).
A Total UF volume set too high can result in an
increased number of LOW DRAIN VOLUME alarms.
(NOTES continued on following page.)
Table 9-17. Hi-Dose Tidal Settings – Low Fill Mode (Continued)
Setting Description