07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 9-13
9. Operating Instructions – Change Program
9.7.3 Hi-Dose CCPD Therapy Settings
Table 9-5 shows the settings programmed for Hi-Dose CCPD therapy in the
Standard Fill Mode.
Table 9-5. Hi-Dose CCPD Settings – Standard Fill Mode
Setting Description
Total Volume
Total Volume of solution used for the therapy. Includes the
total Day Fill Volume and Night Fill Volume for all cycles, and
the Last Fill Volume.
Minimum setting is 200 mL
Maximum setting is 80000 mL
Default setting is 200 mL
Setting increments:
50 in the 200 to 2000 mL range
100 in the 2000 to 5000 mL range
500 in the 5000 to 80000 mL range
# of Day Fills
Number of daytime exchanges. This setting only appears for a
Hi-Dose therapy.
Minimum setting is 0
Maximum setting is 9
Default setting is 0
Setting increment is 1
THERAPY:HI-DOSE CCPD
TOTAL VOL: ML
# OF DAY FILLS: