07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 18-27
18. Troubleshooting
RELOAD THE SET
Display Message:
Cause: The cassette is loaded incorrectly or there is a problem with
the system. See Table 18-1 on page 18-28 for possible
causes.
This is a Manual Restart Alarm.
To Correct: 1. Press
STOP to mute the alarm.
2. CLOSE ALL CLAMPS.
3. Open the door.
4. Remove the cassette.
5. Lift the cassette to drain fluid, if it is full.
6. Reload the cassette.
7. Hold tubing where it exits the cassette holder on your
right and push back while you close the door.
8. Open the clamps.
9. Press
GO.
10. Press
GO again when the display screen shows
CONNECT BAGS.
The priming procedure begins.
RELOAD THE SET