18. Troubleshooting
18-22 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
18.3.11 Caution: Negative UF
WARNING
Bypassing a CAUTION: NEGATIVE UF alarm can leave fluid
in the peritoneal cavity and result in an increased
intraperitoneal volume (IIPV) situation.
IIPV can result in a feeling of abdominal discomfort,
serious injury, or death.
If any patient, or patient caregiver, suspects the patient has
IIPV during a treatment, press
STOP immediately, then
press
V and initiate a Manual Drain. The Manual Drain
procedure is located in 18.5, Manual Drain Procedure, on
page 18-53. See 18.8, Increased Intraperitoneal Volume
(IIPV), on page 18-58 if IIPV is suspected.
Additional care should be taken to monitor for IIPV
symptoms for those patients not able to communicate
essential information to their caregiver during treatment.
CAUTION: NEGATIVE UF
Display Message:
Cause: You have retained more than the allowable percentage of the
programmed Fill Volume in either the current cycle or over
the course of several cycles.
The allowable percentage in Standard Fill Mode is 50%.
In Low Fill Mode, it is adjustable from 20% to 60%.
This is a Manual Restart Alarm.