Red Cycler 5C Instruções de Operação Página 328

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 327
19. Technical Data
19-4 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
In the case of unexplained EMI, consider the locations of nearby
transmitters, such as radio or TV stations. You may have to move the
cycler or place shield material between the transmitter and the cycler.
Be aware that modifying the cycler or adding accessories or components
not specifically authorized by Baxter may make the cycler more
susceptible to interference from radio waves.
The following cables and accessories have been approved for use with the
HomeChoice/HomeChoice PRO APD System, and comply with current EMC
standards:
Detachable power cord
Modem cable
WARNING
When servicing the cycler, use only replacement
components, cables and accessories authorized by Baxter
and be sure to replace all shields, covers, screws, and
gaskets in their exact locations. Failure to do so may result
in increased emissions or decreased immunity of the
cycler.
WARNING
The HomeChoice/HomeChoice PRO APD System should not
be used next to, or stacked with, other electrical
equipment. Such equipment may cause the cycler to
operate incorrectly. However, if it is necessary to use the
cycler close to other equipment, the cycler should be
monitored to verify normal operation.