9. Operating Instructions – Change Program
9-52 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
Dextrose
(default setting)
– or –
The Last Fill Volume can be the same dextrose concentration
as the other Fills or it can be different.
This setting does not appear if LAST FILL VOL = 0.
This setting only appears if you use Last Fill.
Full Drains
Every
This setting appears when the therapy type is set to Tidal, and
Tidal Full Drains is enabled in the Nurse’s Menu by your
nurse.
Minimum setting is 1
Maximum setting is 99
Default setting is 1
Setting increment is 1
h NOTE: The Full Drains Every setting always defaults
to “1” when Tidal Full Drains is enabled. This
must be changed in the Therapy menu to prevent
full drains every cycle.
h NOTE: Use this setting to get a full Drain mid-therapy
when you have a large number of Tidal cycles.
This reduces over- (or under-) fills due to under-
(or over-) estimating the Total UF for the Tidal
therapy.
Table 9-15. Tidal Settings – Low Fill Mode (Continued)
Setting Description
DEXTROSE: SAME
DEXTROSE: DIFFERENT
FULL DRAINS EVERY: