07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 18-55
18. Troubleshooting
18.6 End Therapy Early Procedure
Follow the steps below to end your therapy early.
h
NOTE:
If you end your therapy early for any reason, you can be left with
more fluid in your peritoneal cavity than normal. If this occurs, your
Initial Drain Alarm (I-DRAIN ALARM) setting may be too low. To
minimize the potential for an increased intraperitoneal volume (IIPV)
situation, do one of the following at the beginning of your next
therapy:
If a VERIFY I-DRAIN: ML prompt appears, press STOP and
press
U or V to increase your I-DRAIN ALARM setting to at least
70% of your current expected peritoneal volume for this therapy
only.
– OR –
If a VERIFY I-DRAIN prompt does not appear, press STOP and V
to MANUAL DRAIN. Press
ENTER to initiate a Manual Drain. The
system will return to STOPPED: DRAIN when the Manual Drain
ends. You can repeat the Manual Drain any number of times
without an audible alarm. Resuming the Drain can result in an
audible alarm.
Steps to end therapy early Display screen
1. Press the power switch OFF.
2. Press the power switch ON.
The alarm sounds.
3. Press
STOP.