07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 9-43
9. Operating Instructions – Change Program
Last Fill
Volume
Last Fill Volume delivered at the end of the therapy and left in
the peritoneal cavity during the day. Also called “Wet Day.”
Minimum setting is 0 or 60 mL
Maximum setting is 1000 mL
Default setting is 0 mL
Setting increments:
1 in the 60 to 100 mL range
10 in the 100 to 500 mL range
50 in the 500 to 1000 mL range
Dextrose
(default setting)
– or –
The Last Fill Volume can be the same dextrose concentration
as the other Fills or it can be different.
This setting does not appear if LAST FILL VOL = 0.
This setting only appears if you use Last Fill.
Table 9-13. Hi-Dose CCPD Settings – Low Fill Mode (Continued)
Setting Description
LAST FILL VOL: ML
DEXTROSE: SAME
DEXTROSE: DIFFERENT